GALEX Instrument

bullet

Design/Performance Summary

bullet

Instrument Photo

bulletScience Calibration Sample Data

Design/Performance Summary

Telescope Aperture 50 cm
Optical Design Modified Ritchey-Chretien with Aspheric corrected
Field of View 1.2 degrees, circular
Focal Length 3 m
Telescope coatings Al+MgF2
Imaging/Grism Modes Optics wheel with (1) CaF2 Imaging window, (2) CaF2 transmission grism; (3) Opaque position.
Grism Rotation Grism position angle may be selected with a resolution of 0.3 degrees.
Dichroic/Corrector
bulletAspheric astigmatism corrector
bulletIon-etched fused silica (aspheric surfaces on both sides)
bulletDichroic beam splitter with dielectric multilayer coating on input side
 

FUV Channel

NUV Channel

Beam path Reflected from dichroic Transmitted through dichroic
Filters Blue edge filter (blocks OI, Ly a) - transmission Red block filter/Fold mirror (blocks red Zodaical light)
Detectors Sealed tubes with crossed delay-line anodes
Detector Window MgF2 Fused Silica
Detector Photocathode CsI, opaque CsTe, semitransparent
Detector peak QE 12% 8%
Detector maximum local countrate w/o saturation 100 c/s 100 c/s
System angular resolution 5 arcsec (80% EE diam)

3.3 arcsec (FWHM)

7.5 arcsec (80% EE diam)

5 arcsec (FWHM)

Spectral Resolution ~200 ~100
Imaging Effective Area 25 cm2 44 cm2
Spectroscopy Effective Area 20 cm2 39 cm2

 

Instrument Photo

This photo shows the GALEX instrument in flight configuration (all thermal blankets installed) prior to Thermal Vacuum testing.

 

 

 

 

 

 

 

 

 

 

 

Science Calibration Data

Flat Field Fine Linearity Simulated Sky
Scanned Simulated Sky Relative Sensitivity Grism calibration